LED Driver Testing
What are the limitations of LM-79?
What are the limitations of LM-79?
As we all know about the strength of LM 79. Everything has its pros and cons. Let us have a look on the drawbacks of LM 79.
- LM 79 depicts the parameters of the individual products. So the obtained Parameters cannot be used to access alike products.
- Other limitation is the sample size. It does not state any sample size. So, the manufacturer provides only one sample instead of supplying more. This thus increases the vulnerability of lesser quality products.
- This can be used only to determine complete luminary instead of LED used ( as based on absolute photometry) , which can produce different results due to quality of accessories used like glass, thermal effect, lenses etc.
If you’re Buyers Demanding for LM-79 Testing– Contact Now!
Wish this blog post will be help. If you have any other questions, please feel free to contact us. ITCIndia team looking forward work with you and get better together!
Mr. Puneet Sharma | Call: 08196980555 | E-mail: [email protected]
Get Instant Quotations, Fill the Enquiry form:
This entry was posted in LED Bulb Testing, LED Driver Testing, LED Street Light Testing, LED Testing, LED Tube Light Safety Testing, LM 79 Tesing For Street Light As Per IEC 60598-2-3, LM 79 Testing, LM 79 Testing for LED Floodlights and Streetlights, LM 79 Testing in Kolkata West Bengal India, LM 79 Testing Mumbai Maharashtra India, LM-79 Testing Ahmedabad, LM-79 Testing for LED Batten, Lm-79 Testing for LED COB Down light, Lm-79 Testing for LED COB Spot Lights, Lm-79 Testing for LED COB Zoom light, LM-79 Testing for LED DownLight, LM-79 Testing for LED Flood lights, LM-79 Testing for LED Solar Street Light, Lm-79 Testing for Tube Light, LM-79 Testing in Bangalore Karnataka India, Lm-79 testing lab in India, LM-79 Testing on High bay LED lights, LM-79 testing on High Power LED Light Model, LM-79 Testing | LM-79 Test Approved Methods, Photometric Testing, Photometric Training and tagged & TM21, and TM-21, Calculating SSL Performance from LM-79 -ITCIndia, ISTM, istmt, LED Lighting Facts, led lumen maintenance, led testing labs in india, Lm 79, lm 79 80 standards, lm 79 certificate, lm 79 test, lm 79 testing labs in bangalore, lm 79 testing labs in india, lm 79 testing labs in mumbai, lm 80 report, lm-79 and lm-80 report, LM-79 Luminaire Testing Laboratory, lm-79 requirements, lm-79 test procedure, LM-79 Test Report, LM-79 test reports and luminaire LED quality and performance, LM-79 testing, LM-79 Testing - ITCINDIA, Lm-79 testing Ahmedabad, Lm-79 testing Chennai, Lm-79 testing Delhi, Lm-79 testing India, lm-79 testing kolkata, lm-79 testing lab, Lm-79 testing lab in India, Lm-79 testing Mumbai, lm-80, lm79, lm79 and lm80 standards, lm79 and lm80 testing, lm79 report, lm80 certification, lm80 test procedure, luminaire testing, luminaire testing lab, luminaire testing laboratory, Luminaries/LED Product Testing, sample lm79 report, Searches related to What are the limitations of LM-79, tm 21, Understanding IES LM-79 & IES LM-80, Understanding LED tests IES LM-79, What are the limitations of LM-79?, What are the limitations of LM-79? « Electrical Safety Testing.
LED Driver and Lamp Control Gear Testing
ITCIndia can do LED Driver and Lamp control Gear testing as per IEC 61347 -2-13, IS 15885-2-13 Standards
If you are looking for LED Driver and Lamp control Gear Testing Laboratory for LED Driver and Lamp control Gear In India, Mumbai, Chennai, Ahmedabad, New Delhi, Pune, Kolkata, Bangalore, Hyderabad, Surat, Visakhapatnam and other city, ITCINDIA NABL Approved Electrical Safety testing Lab Can Help You! ITC India one of the best LED Driver and Lamp control Gear testing Laboratory in India.
ITCIndia “Institute of Testing and Certification India Pvt. Ltd” are a leading testing company/laboratory having a NABL accreditation under ELECTRICAL, ELECTRONICS and PHOTOMETRY testing with certificate no. T-3509, T-3570 and T-1811. The Laboratory is following the procedure as required by International Standard IEC 17025 and is already certified for ISO/IEC 17025.
Applicable Standards:
- IEC 61347 -2-13
- IS 15885-2-13
Essential Tests:
- Protection against accidental contact with live parts
- Protective earthing
- Moisture resistance and insulation
- Electric strength
- Endurance test
- Fault condition
- Abnormal conditions
- Creep age and clearance
- Resistance to Heat, fire and tracking
- Resistance to corrosion
Some important table of the issued report ……
Table 1 | |||
Clearances | |||
RMS working voltage not exceeding 250V | Minimum Value (mm) | Measured Value (mm) | Verdict |
– Basic insulation | >1.7mm | >1.7mm | P |
Creepage | |||
RMS working voltage not exceeding 250V | Minimum Value (mm) | Measured Value (mm) | Verdict |
– Basic insulation PTI ≥ 600 | 1.7mm | >1.7mm | P |
Table 2 | ||||
Sr. No. |
Conducted Test |
Test Conditions |
Test requirements |
Result |
1 |
Insulation Resistance Test |
Test Voltage: 500V D.C Time : 1 min Test part: Between live part and metal covered |
Shall not be Less than 2 MΩ |
P (>9900mΩ) |
2 |
Dielectric Strength test |
Test voltage:1480V Test Duration: 1 min. Ramp time: 10sec
Test Part: Between live part and metal covered |
Shall withstand without breakdown |
P |
For your better understanding, we are going to show you the LED Driver and Lamp control Gear Test report of latest LED in the following:
Learn more about this LED Driver and Lamp control Gear test >
If you’re Buyers Demanding for LED Driver and Lamp control Gear Testing– Contact Now!
Wish this blog post will be help. If you have any other questions, please feel free to contact us. ITCIndia team looking forward work with you and get better together!
Mr. Puneet Sharma | Call: 08196980555 | E-mail: [email protected]
Get Instant Quotations, Fill the Enquiry form:
This entry was posted in Lamp Control Gear Testing, LED Driver Testing, LED Testing and tagged 16104 : 2012, 16105 : 2012, Ahmedabad, Bangalore, Chennai, electronic control gear for led modules, How do I test a led driver, Hyderabad, iec 61347-1 pdf, iec 61347-2-13, is 15885 part 2 sec 13, is 15885 part 2 sec 3, is 15885 part 2 section 13, IS 15885-2-13, Kolkata, Lamp Control Gear Testing, Lamp control Gear Testing as per IEC 61347 -2-13, Lamp control Gear Testing as per IS 15885-2-13, LED Driver and Lamp control Gear In India, LED Driver IS 15885-2-13, LED Driver Testing, LED Driver Testing - ITCINDIA Lab, LED Driver Testing as per IEC 61347 -2-13, led driver testing equipment india, led driver testing instruments, led driver testing method, led driver testing standards, led driver troubleshooting, led light testing procedure, LED Power Driver Test Solution, LED Testing - Electrical Safety, lt-101a led driver tester, Mumbai, New Delhi, Pune, Searches related to Lamp control Gear Testing, Searches related to LED Driver testing, Surat, Test of LED Driver, testing led driver with multimeter, Testing LED Drivers, Visakhapatnam.